pue41Towary_id2kod 0
Setting value 1 will send to PUE_C41 scale ID
from the database as the scale code;
The name of product on the scale = the
name of product from the computer
The code of product on the scale = ID
name of product from the computer
It is used when the operator’s code is longer
than 6 characters; the operator’s code for the
scale cannot be longer than 6 characters.
pue41Towary_id2kod
_MixName
0
Setting value 1 will send PUE_C41 to the
scale and use on the scale the ID from the
database as a code on the scale together with
the name change for the name with the ID at
the beginning
The product on the scale name = ID of
product from the computer + name of the
product from the computer
Code of the product on the scale = ID of
product from the computer (the code
cannot be longer than 6 characters)
Used when the products names are identical
on the computer; the products names on the
scale must be unique
The option works if you set
up 1 for
pue41Towary_id2kod.
Synch_Towar
MasaEnabled
0
Setting the parametre on value 1 will turn on
update of the mass standard from all scales
Synch_Towar
MinMaxEnabled
0
Setting the parametre on value 1 will turn on
update of thresholds from all scales
Synch_Towar
TaraEnabled
0
Setting the parametre on value 1 will turn on
update of the tare value from all scales
Synch_Towar
UpdateEnabled
0
Setting the parametre on value 1 will turn on
the update service from all scales to the
database