User manual no.:
IMMU-12-03-04-14-ENG
MOISTURE
ANALYZER
MAC
SERIES
BALANCES AND SCALES
RADWAG 26 600 Radom, Bracka 28, POLAND
Phone: +48 (0-48) 38 48 800, fax. +48 (0-48) 385 00 10
export@radwag.com
www.radwag.com
Przeglądanie stron 0
1 2 3 4 5 6 ... 59 60

Podsumowanie treści

Strona 1 - ANALYZER

User manual no.: IMMU-12-03-04-14-ENG MOISTURE ANALYZER MAC SERIES BALANCES AND SCALES RADWAG 26 – 600 Radom, Bracka 2

Strona 2

10 considerably shortens measurement time and simplifies testing procedure. MA 3Y series allows for setting multiple parameters which influence the

Strona 3

11 with the supply in the mains to which the moisture analyzer will be plugged in, - replecemant of halogen lamps can only be carried out by autho

Strona 4

12 2.9.1. Protective clothing The use of protective clothing is highly recommended while working with the instrument in order to take safety preca

Strona 5

13 • Ambient relative humidity should not exceed 80% in the temperature up to 31°C, and decrease linearly to 50 % of relative humidity in temperatu

Strona 6

14 When assembling the pan supporter (bracket) pay special attention to the correct positioning of its mandrel. The mandrel features a cut for it

Strona 7

15 4.6. Plugging to mains A moisture analyzer can be plugged to mains only by means of original power cord, which comes standard with the moisture

Strona 8

16 5. KEYBOARD – FUNCTION KEYS Start/Stop key – starting / ending drying process according to selected drying mode. Esc key, abandoning

Strona 9

17 Display key – changes type of data displayed during and after drying process. Setup key – enter to main menu. Test menu key – function key f

Strona 10

18 7. USER MENU User menu is divided into 5 basic function groups. 7.1. Moving through user menu An operator moves through moisture analyzer’s men

Strona 11

19 Names of function groups and their content: P1 CAL [Adjustment] P1-01 ECAL | [external adjustment] P1-02 tCAL | [adjustment test] P1-

Strona 12

2 MARCH 2014

Strona 14

21 P6 othEr [Other functions] P6-01 Libr | YES/no [access to drying modes library] P6-02 bL | YES/no [display backlight] P6-03 bLbA |

Strona 15

22 8. WEIGHING Before start of measuring process or in case of essential change of ambient conditions at a workstation (e.g. ambient temperature c

Strona 16

23 User of a moisture analyzer can adjust its operation to current ambient conditions at a workstation (by changing the value of digital filters), o

Strona 17

24 CAUTION: The stronger filtering setting the longer weighing time. 9.2. Autozero function In order to ensure moisture analyzer’s precise mass indi

Strona 18

25 9.3. Access to library of drying programs In order to use the library of drying programs set parameter <Libr> to value <YES>: no

Strona 19

26 9.4. Backlight of measurement result (display backlight) The function enables switching on and off backlight of measurement result on the displa

Strona 20

27 9.5. Brightness adjustment of display backlight 100 - maximum display brightness 20 - minimum display brightness no - display brightne

Strona 21

28 10. MOISTURE ANALYZER ADJUSTMENT As the gravitational acceleration force has different value at various latitudes and altitudes, a moisture anal

Strona 22

29 10.1. External adjustment External adjustment in moisture analyzer MAC series should be carried out with an external mass standard / weight cla

Strona 23

3 1. TECHNICAL DATA ... 5 2. GENERAL SAFETY INFORMATION ...

Strona 24

30 10.2. Adjustment test Adjustment test is a comparison of mass of an adjustment weight with the value of last adjustment saved in moisture analy

Strona 25

31 10.3. Drying chamber adjustment Special adjustment set, which is an atypical accessory of a moisture analyzer, is used for temperature adjustmen

Strona 26

32 Step 2. Insert parts of the set into the drying chamber: • shield and thermometer Step 3. After set assembly close the drying chamber and st

Strona 27

33 Go to menu P1-03 tE_CAL, adjustment process is initiated automatically. The moisture analyzer displays current temperature in the drying ch

Strona 28

34 to the current temperature displayed on the control thermometer – as in previous step. On entering the temperature value and accepting it by pr

Strona 29

35 Remember, that if the parameter is set to YES, then a report is generated and sent (printed) automatically. Content of report from adjustment pro

Strona 30

36 Enter P1 CAL menu, next start P1-05 tSt_Co procedure. Before the test activation set its parameters in accordance with diagram and description,

Strona 31

37 After 8 minutes the user will see a window, where temperature read from the assembled adjustment set must be entered. Use navigating arrows t

Strona 32

38 • P2-01 USr The parameter enables determining name of a operator who works with a moisture analyzer. User name contains maximally 6 alphanumeri

Strona 33

39 WILK An instance of a user name inserted to a moisture analyzer using small letters takes below form: wilk • P2-02 Prj The parameter enable

Strona 34

4 11. DETERMINING CONTENT OF A PRINTOUT FOR GLP PROCEDURES ... 37 12. D

Strona 35

40 Option determining presence of moisture analyzer’s factory number on a printout. • P2-08 PFr Option enabling printing frames on a printout. F

Strona 36

41 12.2. Setting date This parameter enables setting date in a moisture analyzer. In order to change date settings use navigating keys Up, Down,

Strona 37

42 - 9600 bit/s - 19200 bit/s. Return to weighing mode (see point 7.2 – return to weighing mode). 13.2. Determining data type sent via RS 2

Strona 38

43 13.3. Printout interval Printout interval determines how often data on drying process is sent through RS 232 port. Available settings: 1, 2,

Strona 39

44 Return to weighing mode (see point 8.2 – return to weighing mode). 14. DRYING PROCESS WITH USE OF SHORTENED MENU Menu of drying mode enable

Strona 40

45 14.1. Setting means of auto switch off mode Parameter switch off mode can take the following values: 1 – automatic finish (mass change 1

Strona 41

46 In order to accept carried out changes, press Print/Enter key for three (3) times. 14.2. Active drying mode

Strona 42

47 In order to accept carried out changes, press Print/Enter key for three (3) times. Parameter active drying mode can take the following values:

Strona 43

48 • Drying mode STEP In step mode, user should set drying temperature tmP in which a sample should be tested, temperature values tmP1 and t

Strona 44

49 CAUTION: For drying mode Quick the user must set the drying temperature and time interval of heating in temperature 30% higher than the specif

Strona 45

5 1. TECHNICAL DATA Type MAC 50/1 MAC 50 MAC 110 MAC 210 Max capacity 50 g 50 g 110 g 210 g Readability 0,1 mg 1 mg 1 mg 1 mg Tare range -

Strona 46

50 Displays mass change recorded during drying process, and expressed in percent, • Part of dry mass obtained in drying process, and expressed

Strona 47

51 In order to exit the option of using drying program libraries, disable access to libraries in moisture analyzer’s menu (see point 9.3. of this

Strona 48

52 16. DRYING PROCESS In order to start the drying process, a user should select a drying mode, and then act according to below guidelines:

Strona 49

53 Start drying process by pressing Start/Stop key, and on emptying moisture analyzer’s drying pan, pressing TARE key. Open the drying chamber’s

Strona 50

54 Changing of displayed data during drying process is carried out by pressing Display key.

Strona 51

55 In order to abort drying process, press Start/Stop key and accept switch off mode of drying process by pressing Print/Enter key. An excepti

Strona 52 - 16. DRYING PROCESS

56 17. SET OF COMPUTER – MOISTURE ANALYZER COMMANDS Function INTERFACE RESET Command R CR LF (zeroing of currently carried out commands, i.e.

Strona 53

57 Function IMMEDIATELY SEND THE RESULT IN CURRENT UNIT Command SUI CR LF (sends the result in current used unit from the instrument) Function

Strona 54 - Display key

58 18. ERROR MESSAGES If moisture balance software detects an error a proper error message appears on the graphic display. The error message conta

Strona 55

59 Parameter allowing the moisture analyser to be operated by means of a prior version of the software (valid since June 2013). Upon setting paramet

Strona 56

6 Dimensions: 2. GENERAL SAFETY INFORMATION 2.1. Definitions of signals and warning signs Safety precautions are marked with special descript

Strona 57

60 MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG BALANCES AND SCALES POLAND, 26 – 600 Radom, Bracka 28 Phone. +48 48 38 48 80

Strona 58

7 Acid / Corrosion Danger Flammable or explosive substances Toxic substances Hot surfaces 2.2. Safety information CAUTION! The use

Strona 59

8 forbidden. CAUTION: Drying chamber cannot be opened during operation (drying process). It is due to the fact that a halogen lamp and its glas

Strona 60

9 Flammable or explosive substances, substances containing solvents or releasing flammable or explosive gases or vapours. Perform a risk analysis