15.1. Calibration
Procedure:
Enter the submenu <P6.CAL> and then:
Following inscriptions will appear
A new start mass is adjusted during this period of time.
After that a mass of calibration weight is shown (e.g. 3 000kg).
Put a weight of the displayed mass value on the pan and press
.
The calibration process will start which is signalled by the message:
After completion of the process of calibration the following screen
will appear