measuring range of a balance (the <TARE>
button applied), as well as on weighing
samples without the packaging (the
<TARE> button is not applied).
Example no. 3 a balance AS 220.3Y:
The minimum sample refers to all net
weighments that are weighed without a
packaging (the <TARE> button is not
applied).
The software identifies this setting as the
minimum sample is valid only for weighing
samples without a packaging. If tarring
option is applied then the software disables
the pictogram informing on minimum
sample in use, and renders the weighment
as carried out with tare applied, that does
not have a minimum sample criterion
The user can preview entered data, but data editing is disabled.
27.13. Ambient conditions
Database of ambient conditions contains parameters related to measurement
of ambient conditions at a workstation. Depending on balance’s configuration,
a report may contain data on temperature, humidity and value of atmospheric
pressure. In case a THB ambient conditions module is connected to a balance,
records from the module are also saved.
Procedure
• Enter submenu < Databases>, and press < Ambient conditions >
field
• Press field of a desired report record, unless it is visible on the top of the
list, wind the list forward using navigating keys
• Report’s name consists of date and time
Caution:
It is possible to use an option for searching a report.
27.14. Packages
Database of packages contains list of packages for which a name, code and
mass value are specified. During weighing process, if a name of packaging is
selected from the database, the tare value is automatically applied. The
display indicates mass value with minus sign.