Filter Very fast
Value release Reliable
Repeatability 0.0092 g
Stabilization time 4.895 s
Filter Fast
Value release Fast
Repeatability
0.0099 g
Stabilization time 4.979 s
.
.
.
Filter Very slow
Value release Reliable
Repeatability 0.0207 g
Stabilization time 5.015 s
Signature
........................................
PROCEDURE:
On function activation, the software initiates the process automatically. The
display indicates progress of the process course. On completing the process, the
software displays a summary and marked values of currently set filter values.
The report can be printed.
Caution:
Exiting the reporting window causes erasing the test results.
User can abort the process at optional moment by pressing <X> key in the
testing window.
AUTOTEST GLP
The test controls repeatability of placing the internal weight and
determining error of indication with reference to balance’s Max capacity.
Testing Procedure:
Two times loading of the internal weight, followed by 10 times
loading of the internal weight
Balance internal adjustment
Calculating and storing the value of standard deviation
In balances with automatically opened doors, door opening and
closing test is carried out.
Additionally the function enables:
Displaying and printing a report on available printers and saving
the report in balance’s memory. The report contains basic data on
balance’s parameters, ambient conditions and test results.