Radwag MYA 5.3Y.F1 Microbalance Instrukcja Użytkownika Strona 177

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 198
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 176
- 177 -
Example1
Command:
OMI <CR><LF> give available working modes
Response:
OMI <CR><LF> available working modes are given
2_” Parts counting” <CR><LF> number + name
4_” Dosing” <CR><LF>
12_”Checkweighing” <CR><LF>
OK <CR><LF> command completed
Example2
Command:
OMI <CR><LF> give available working modes
Response:
OMI <CR><LF> available working modes are given
2 <CR><LF> (mode number)
4 <CR><LF>
12 <CR><LF>
OK <CR><LF> command completed
OMS - Set working mode
Command sets current working mode of a balance.
Format: OMS_n <CR><LF>
Accessible responses:
OMS_OK <CR><LF>
command completed
OMS_E <CR><LF>
error occurance in course of command
performance, incorrect format or lack of the
parameter
OMS_I <CR><LF>
command understood but not accessible at this
moment
n parameter, decimal value specifying working mode number. See OMI
command description for detailed information.
Example
Command:
OMS_13<CR><LF> set Statistics mode
Response:
OMS_OK<CR><LF> Statistics mode selected
OMG - give current working mode
Command gives current working mode for a given device.
Format: OMG <CR><LF>
Accessible responses:
OMG_n_OK
command completed, current working mode is
Przeglądanie stron 176
1 2 ... 172 173 174 175 176 177 178 179 180 181 182 ... 197 198

Komentarze do niniejszej Instrukcji

Brak uwag